SAT – Saturation with variable limits
Block SymbolLicensing group: STANDARD
Function Description
The SAT block copies the input u to the output y if the input signal satisfies
and
,
where lolim and hilim are state variables of the block. If
(),
then
(). The upper and lower
limits are either constants ()
defined by parameters hilim0 and lolim0 respectively or input-driven variables
(, hi and lo
inputs). The maximal rate at which the active limits may vary is given by time constants tp (positive
slope) and tn (negative slope). These rates are active even if the saturation limits are changed
manually ()
using the hilim0 and lolim0 parameters. To allow immediate changes of the saturation limits,
set
and .
The HL and LL outputs indicate the upper and lower saturation respectively.
If necessary, the hilim0 and lolim0 parameters are used as initial values for the input-driven saturation limits.
Inputs
u | Analog input of the block | Double (F64) |
hi | Upper limit of the output signal (for the case ) | Double (F64) |
lo | Lower limit of the output signal (for the case ) | Double (F64) |
Outputs
y | Analog output of the block | Double (F64) |
HL | Upper limit saturation indicator | Bool |
LL | Lower limit saturation indicator | Bool |
Parameters
tp | Time constant defining the maximal positive slope of active limit changes 1.0 | Double (F64) |
tn | Time constant defining the maximum negative slope of active limit changes 1.0 | Double (F64) |
hilim0 | Upper limit of the output (valid for ) 1.0 | Double (F64) |
lolim0 | Lower limit of the output (valid for ) -1.0 | Double (F64) |
HLD | Fixed saturation limits on | Bool |
|
|
|
[Previous] [Back to top] [Up] [Next]
2024 © REX Controls s.r.o., www.rexygen.com